Как выбрать гостиницу для кошек
14 декабря, 2021
Sensor manufacturers subject their designs to a series of tests simulating actual operating conditions to determine the on-line operating characteristics. Standard definitions are given in the Scientific Apparatus Makers Association (SAMA) publication PMC-20, Measurement and Control Terminology.
A sample performance report on a motion-balance
sensor is given below
Description
Pressure-sensing mechanism Bourdon tube (316 stainless steel) Electric transmission Output-signal ranges
±10 volts d-c, ±50 mV d-c, 0 to 100 mV d-c Operating Conditions
Ambient temp, nominal, 75°F, reference, calibration ±5°Г, normal, 40 to 140° F, operative limits, -10 to 200° F Supply voltage nominal, 118 volts a-c, normal, 107 to 127 volts, operative limits, 100 to 1 35 volts Frequency nominal, 50 or 60 H7, normal, 48 to 62 Hz, operative limits, 45 to 75 Н/
Ambient temp, effect 7ero-shift error/100°F temp, change, — 1% range span, Range-shift error/100°F temp change + 1% range span
Reference Performance Characteristics (°0 range span)
Accuracy 0.5%
Dead band 0.2%
Hysteresis 0.5%
Linearity 0.25%
Repeatability 0.25%
Design Data
Source impedance a-c signal coil, 200 ohms, d-c signal demodulator, 180 ohms
Minimum external load a-c transmitted signal, 2000 ohms, d-c transmitted signal, 30,000 ohms Maximum ripple 0.15% a-c ripple
Case classification NEMA (National Electrical Manufacturers Association) type 2 or N1 MA type 7D Over-range protection ll/4 times max. scale measured pressure
Performance data on a force balance sensor is given as
Description
Pressure-sensing capsule 316 stainless steel Electric transmission 2 wire d-c Output signal range 10 to 50 mA d-c Operating Conditions
Power supply 63 to 85 volts d-c Supply voltage effect 0 25% per 10-volt variation Performance characteristics (% range span)
Accuracy 0 5%
Dead band 0.005%
Repeatability 0.15%
Design Data
Output load limits 600 ohms (+10%, -20%)
Case classification NEMA type 4, hazardous area Class I Group D, Div. 1
(a) Pressure Switches. These are widely used to actuate alarms or initiate sequential operations. A Bourdon tube or similar sensor is linked to a snap-acting mechanical switch.
(In some cases an enclosed mercury switch is used.) The switch may be indicating or nomndicating, have rangesetting capability, and provide necessary logic at predetermined pressures.
(b) Electric Modulating Transmitters. These produce an electrical output proportional to input pressure (or force) applied to the sensor. Either the motion-balance or the force-balance principle may be involved. The output may be a voltage or a current of suitable value and range for input to readout devices, such as recorders, indicators, computers, and control loops to action equipment. A sample circuit for the motion-balance example of Sec. 4-3.6 is shown in Fig. 4.23. Forms of the linear voltage differential transformer (LVDT) mechanism and a sample output curve are shown in Fig. 4.24.
(c) Pneumatic Modulating Transmitters. Differential — pressure sensors installed with one side open to the atmosphere and the other side connected to a pressure source can be used The device shown in Fig. 4.35 can be used and the pneumatic force-balance principle applied to obtain a pneumatic output proportional to sensor gage pressures at connection H (or L, as desired).